MECHANISM
|
TECHICAL DESCRIPTION
|
01
|
Channel
|
Dual, isolated between channels |
02
|
Pulse Amplitude
|
Adjustable 0-90 mA into 500 ohm load each channel. |
03
|
Wave Form
|
Asymmetrical Bi-Phasic Square Pulse |
04
|
Voltage
|
0 to 45V (Load: 500 ohm) |
05
|
Power source
|
One 9 Volt Battery. |
06
|
Size
|
10.1cm(L) x 6.1cm(W) x 2.45cm(H) |
07
|
Weight
|
150 grams with battery. |
08
|
Pulse Rate
|
Adjustable, from 2 to 150 Hz, 1 Hz/step |
09
|
Pulse Width
|
Adjustable, from 50 to 300 microseconds, 10 µs/step |
10
|
On Time
|
Adjustable, 2~90 seconds , 1 Sec./ step |
11
|
Off Time
|
Adjustable, 2~90 seconds , 1 Sec./ step |
12
|
Ramp Time
|
Adjustable, 1~8 seconds, 1 Sec./ step, The “On”time will increase and decrease in the setting value. |
13
|
Mode
|
Five TENS Modes: B(Burst), N(Normal),M (Modulation),SD1( Strength Duration), SD2
Three EMS Modes: C(Constant)
S (Synchronous), A (Alternate) |
14
|
Burst Mode
|
Burst rate: Adjustable, 0.5 – 5Hz
Pulse width adjustable, 50~300µs
Frequency fixed = 100 Hz |
15
|
Normal Mode
|
The pulse rate and pulse width are adjustable. It generates continuous stimulation based on the setting value. |
16
|
Modulation Mode
|
Modulation mode is a combination of pulse rate and pulse width modulation. The pulse rate and width are automatically varied in a cycle pattern. The pulse width is decreased by 50% from its original setting in 0.5 second, then the pulse rate is decreased by 50% from its original setting in 0.5 second. Total cycle time is 1 second. In this mode, pulse rate(2-150Hz) and pulse width(50-300µs) are fully adjustable. |
17
|
SD1 Mode
|
The SD1(Strength-Duration) mode consists of automatic modulation intensity and pulse width in 40% range. The intensity is always increasing while the pulse width is decreasing and vice-versa. The intensity is decreased by 40% while the pulse width is increased by 40% in 5 seconds. In the next 5 seconds, the intensity is increased by 40% while the pulse width is decreased by 40%. Total cycle time is 10 seconds. Pulse rate ( 2~150Hz) and pulse width (50~300µs) are
fully adjustable. |
18
|
SD2 Mode
|
The SD2(Strength-Duration) mode consists of automatic modulation intensity and pulse width in 70% range. The intensity is always increasing while the pulse width is decreasing and vice-versa. The intensity is decreased by 70% while the pulse width is increased by 70% in 5 seconds. In the next 5 seconds, the intensity is increased by 70% while the pulse width is decreased by 70%. Total cycle time is 10 seconds. Pulse rate ( 2~150Hz) and pulse width (50~300µs) are
fully adjustable.
|
19
|
Constant Mode(C)
|
The pulse rate and pulse width are adjustable. It generates continuous stimulation based on the setting value.
|
20
|
Synchronous Mode(S)
|
Stimulation of both channels occurs
Mode(S) synchronously. The “ON” time including “Ramp Up”and “Ramp Down” time. Therefore, the setting of ON Time should be no less than two times of the “Ramp”time in this mode.
ON TIME ≥Ramp up + Ramp down |
21
|
Alternate Mode(A)
|
The stimulation of the CH2 will occur after the 1st contraction of CH1 is completed. In this mode, the setting of ON Time should be no less than two times of the “Ramp” time. The OFF Time should be equal or more than the ON Time.
ON TIME ≥Ramp up + Ramp down
OFF TIME ≥ IN TIME
|
22
|
Timer
|
Selectable, 15, 30, 45, 60 minutes and C (Continue) |
23
|
Patient Compliance Meter
|
The accumulated treatment time can be recorded. Total recorded time is 999 hours
|
24
|
Constant Mode(C)
|
Temperature:0°~40°C
Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1060Hpa
|
25
|
Remark
|
There may be up to a +/-10% tolerance of all parameters and +/-20% tolerance of output amplitude & voltage.. |